Limnology and Oceanography: Methods, Volume 20, Issue 4


Anthology ID:
G22-23
Month:
Year:
2022
Address:
Venue:
GWF
SIG:
Publisher:
Wiley
URL:
https://gwf-uwaterloo.github.io/gwf-publications/G22-23
DOI:
Bib Export formats:
BibTeX MODS XML EndNote

pdf bib
Improvement of field fluorometry estimates of chlorophyll <i>a</i> concentration in a cyanobacteria‐rich eutrophic lake
Amir M. Chegoonian | Kiana Zolfaghari | Peter R. Leavitt | Helen M. Baulch | Claude Duguay

Instrumented buoys are used to monitor water quality, yet there remains a need to evaluate whether in vivo fluorometric measures of chlorophyll a (Chl a) produce accurate estimates of phytoplankton abundance. Here, 6 years (2014–2019) of in vitro measurements of Chl a by spectrophotometry were compared with coeval estimates from buoy-based fluorescence measurements in eutrophic Buffalo Pound Lake, Saskatchewan, Canada. Analysis revealed that fluorometric and in vitro estimates of Chl a differed both in terms of absolute concentration and patterns of relative change through time. Three models were developed to improve agreement between metrics of Chl a concentration, including two based on Chl a and phycocyanin (PC) fluorescence and one based on multiple linear regressions with measured environmental conditions. All models were examined in terms of two performance metrics; accuracy (lowest error) and reliability (% fit within confidence intervals). The model based on PC fluorescence was most accurate (error = 35%), whereas that using environmental factors was most reliable (89% within 3σ of mean). Models were also evaluated on their ability to produce spatial maps of Chl a using remotely sensed imagery. Here, newly developed models significantly improved system performance with a 30% decrease in Chl a errors and a twofold increase in the range of reconstructed Chl a values. Superiority of the PC model likely reflected high cyanobacterial abundance, as well as the excitation–emission wavelength configuration of fluorometers. Our findings suggest that a PC fluorometer, used alone or in combination with environmental measurements, performs better than a single-excitation-band Chl a fluorometer in estimating Chl a content in highly eutrophic waters.